5.11 Menu 11 - Functions
Menu
Position
Converter Function Code Function Name Function Description
Access
Level
MCP
Command
11.1
[T+ reset]
Totalizer Total
Positive reset
Resets the total direct totalizer for
direct ow rate (+)
AL3 [VTTPR]
11.2
[P+ reset]
Totalizer Par.
Positive reset
Resets the total partial totalizer for
direct ow rate (+)
AL3 [VTPPR]
11.3
[T- reset]
Totalizer Total
Neg. reset
Resets the total reverse totalizer for
direct ow rate (-)
AL3 [VTTNR]
11.4
[P- reset]
Totalizer Partial
Neg. reset
Resets the partial reverse totalizer
for direct ow rate (-)
AL2 [VTPNR]
11.5
[Load Sens.f.def]
Load factory
default sensor
Resets the parameters of the sensor
factory default
AL3 [LFDSD]
11.6
[Load Conv.f.def]
Load factory
default converter
Resets the parameters of the
converter factory default.
AL3 [LFDCD]
FUNCTION: ONLY MCP FUNCTIONS
Menu
Position
Converter Function Code Function Name Function Description
Access
Level
MCP
Command
[MCP ONLY]
Saves sensor reference data AL4 [SRFDS]
5.12 Menu 12 - Diagnostic
Menu
Position
Converter Function Code Function Name Function Description
Access
Level
MCP
Command
12.1
[Self test=]
Self Test
Diagnostic
Meter auto-test function. This
function stops the normal
functions of the meter and
performs a complete test cycle on
the input circuits measurement
and on the excitation generator.
AL3 [ATSIC]
To activate, after selecting the function, press Enter at the question: “CONFIRM EXEC.?” Press ESCAPE to start
the auto-test or any other key to stop the operation. At the end of operation, the converter will revert to one
of the initial visualization pages. This function is automatically performed when switching on the device.
This function restart the converter.
12.2
[Display test=]
Test Display
This function performs a physical
test graphic display. During this
operation, four sequences are
displayed to test that the device
functions properly.
AL3
NO MCP
COMMAND
12.3
[Sens.verify]
Sensor Verify
This function performs a manual
sensor verication ( if BIV is active)
AL3 [SVERC]
30124-60 Rev. 1.0 | unreleased