EasyManua.ls Logo

METER AQUALAB 3 - Page 79

Default Icon
90 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
75
SERVICE
Figure92 Instrument online and visible to NEXUS
5. Follow the directions on the Clean Device tab to make sure the sample drawer and
sensor block are clean (Section4.2).
Figure93 Verify correct instrument
6. Select the Next button.
The standard tab will appear (Figure94). The tab heading will show the selected
standard for the verification test.
If an error message indicates there are no assigned standards for the instrument, go to
skalacontrol.com and select a standard to use.

Other manuals for METER AQUALAB 3

Related product manuals