7 Various functions
780/781 pH/Ion Meter, Manual
95
V total
This is the initial volume of the solution as defined under PARAMETER
Conc/calculation parameters
(see Section 6.5.4).
conc.std
The concentration of the standard used is shown here as defined under
PARAMETER Conc/standard addition.
initial voltage
The initial potential measured in the solution is shown here.
factor
smpl.size
These are further calculation parameters defined under
PARAMETER
Conc/calculation parameters
and used to calculate the final result
(see Section 6.5.4).
add1, add2...addn
For each addition the volume increment, concentration and measured
potential are shown.
curve <> >
The calibration function can be viewed as a plot. The graphics display
is accessed with <
>:
The first measuring point of the solution is marked by '
' and the addi-
tions with a '
+
'.
This display is exited with <
QUIT
>.