8.1 Troubleshooting
114
780/781 pH/Ion Meter, Manual
Problem Possible cause Remedy
Measuring signal absent
or very variable
Electrode not connected
Connect electrode
Air in or in front of the dia-
phragm
Remove air
Faulty electrode
Replace electrode
Measured value remains
unstable and does not
fulfill drift criterion
Dirty pH or ISE membrane or
diaphragm
Clean membrane or diaphragm
pH or temperature of the solu-
tion is not stable
Measure under exclusion of air
Thermostat the solution
Unsuitable electrode:
Conductivity too low
Organic solution
Use suitable electrode
Electrode not connected
Connect electrode
Faulty electrode
Replace electrode
Measured value reached
slowly
Dirty pH or ISE membrane or
diaphragm
Clean membrane or diaphragm
Slope too small on cali-
bration
pH or ISE membrane or dia-
phragm is dirty
Clean membrane or diaphragm
Glass membrane dehydrated
after measurement in non-
aqueous solution
Immerse electrode in water between
measurements
Poor buffer/standard solutions
Replace buffers or standards
Electrode "used up"
Replace electrode
Measured value obvious-
ly incorrect
Incorrect calibration
Check / Repeat calibration
Check / Replace buffer/standard so-
lutions
Manual temperature input incor-
rect
Enter correct temperature of the solu-
tion
Dirty pH or ISE membrane or
diaphragm
Clean membrane or diaphragm
Electrolyte or electrode too old
Refill electrolyte or replace electrode
Faulty electrode
Replace electrode