2.3 Connecting optional devices
14
780/781 pH/Ion Meter, Manual
Printer Cable 780/781 settings Printer settings
Custom
DP40-S4N
6.2134.110 Character Set: Citizen
Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs
IDP-560 EMULATION
FONT MAP = ENGLAND *
PRINT = REVERSE
LITTLE
CR CODE = VOID
CR AFTER B. FULL = VOID
CR ON B. EMPTY = VALID
BUFFER 1K BYTE
BAUD RATE = 9600
PROTOCOL = 8,N,1
FLOW CONTROL CTS-RTS
(*) FRANCE also for Italian, Spanish
and Portuguese,
GERMANY also for Swedish.
Seiko
DPU-414
6.2134.110
Character Set: Seiko
Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs
none
Citizen
iDP562 RS
6.2134.050 Character Set: Citizen
Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs
Epson LX-
300+
6.2134.050 As above, but
Character Set: Epson
see Printer Manual
HP Desk Jet
with serial
interface
6.2134.050 Character Set: HP
Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs
HP Desk Jet
with parallel
interface
2.145.0330
RS232/
Parallel
Converter
Character Set: HP
Baud Rate: 9600
Data Bit: 8
Stop Bit: 1
Parity: none
Handshake: HWs
see Printer Manual
ON
1
2
3
4
5
6
7
8
9
10
SSW1
1
2
3
4
5
6
7
8
1
2
3
4
5
6
7
8
B:
A4 Paper
A:

Table of Contents