3 Short operating tutorial
780/781 pH/Ion Meter, Manual
21
4
Poor calibration data
• If the calibration data are outside the calibration parameters
defined as the limits (see Section 6.2.2) then a corresponding
message will be shown:
• You can still accept the calibration data with
or re-
ject it with
.
4
Show calibration data
• You can now display the calibration data with
.
If you move the selection bar to the end of the list with <
>,
then in the line
Curve
you can activate the calibration curve display with
.
4
Print calibration report
• With
you can now output a calibration report to a
printer or a PC.
• With
you can jump directly to the short calibra-
tion report in the report selection list (cali
all the data except the calibration curve itself.
• The Select key
offers the further option 'calib full';
this also outputs the calibration curve.