3 Short operating tutorial
780/781 pH/Ion Meter, Manual
21
4
Poor calibration data
If the calibration data are outside the calibration parameters
defined as the limits (see Section 6.2.2) then a corresponding
message will be shown:
You can still accept the calibration data with
<ENTER>,
or re-
ject it with
<MODE>
.
4
Show calibration data
You can now display the calibration data with
<CAL.DATA>
.
If you move the selection bar to the end of the list with <
>,
then in the line
Curve
< -> >
you can activate the calibration curve display with
<->>
.
4
Print calibration report
With
<REPORT>
you can now output a calibration report to a
printer or a PC.
With
<CAL.DATA>
you can jump directly to the short calibra-
tion report in the report selection list (cali
b short); this contains
all the data except the calibration curve itself.
The Select key
<->>
offers the further option 'calib full';
this also outputs the calibration curve.
ENTER
MODE
CAL. DATA
7
REPORT
9