XTreme-EMS Operating Manual Page 35
Pressure @ 20mA:
Enter the pressure specified at 20mA output.
Menu options for all sender types
If the “USER" pressure sender is selected
CALIBRATE SENDER:
If the sender type is set to “USER”, then use this menu option to calibrate your pressure sender. See section 4.4.9 for
more information.
PRESSURE UNIT:
Select whether you want to display the pressure in Bar, PSI or PSI(0.1). The PSI(0.1) is for low range pressure senders
e.g. UMA 7PSI.
DISPLAY MAX:
Select the maximum pressure that you want the oil pressure bargraph to show. This can give you increased display
resolution.
DISPLAY MIN:
Select the minimum pressure that you want the oil pressure bargraph to . This can give you increased display resolution.
HIGH ALARM:
This enables or disables the pressure high alarm.
HIGH ALARM:
Enter the pressure threshold for when the high alarm must be activated. Any pressure above this value will activate the
alarm.
HIGH CAUTION:
Enter the pressure value for the high caution. This is the lower value of the upper yellow band.
LOW CAUTION:
Enter the pressure value for the low caution. This is the upper value of the lower yellow band.
LOW ALARM:
This enables or disables the pressure low alarm.
LOW ALARM:
Enter the pressure threshold for when the low alarm must be activated. Any pressure below this value will activate the
alarm.