Page 58
Error Codes
scanCONTROL 30xx
Flashing
sequence
Cause Remedy Notes
--..
2x long, 3x short
Data overflow during transmission
of the data via Ethernet
Reduce profile frequency, increase
packet size.
Data can be faulty.
--…..
2x long, 5x short
Error during calculation
Reduce profile frequency, select faster
calculation mode.
Data can be faulty.
--......
2x long, 6x short
Error during Ethernet transmission Reduce profile frequency. Data can be faulty.
Group: Output Unit
---
3x long
2D/3D Output Unit not found.
Connect the sensor with the 2D/3D
Output Unit.
---
---.
3x long, 1x short
Connected modules of 2D/3D Out-
put Unit are not supported.
Use the supported modules only, see
Chap. 7.4.2.
---
---..
3x long, 2x short
Communication error (2D/3D
Output Unit)
Reduce profile frequency. ---
---...
3x long, 3x short
Output overflow (2D/3D Output
Unit)
Reduce profile frequency. ---
Group: Ethernet interface
----
4x long
IP Address conflict
Check the Ethernet configuration of
device and the host PC.
Choose another IP address for the
device.
If the problem persists, please
contact the manufacturer.
The Data LED flashes green; long during active data transmission and short for controller accesses. A controller access can cause
various data overflows particularly if the measuring frequency is near its maximum.