Functional Tests Periodic Maintenance
3 - 12 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 3-10 Automatic Circuit Leak and Compliance Test: Fail
The following screen is displayed if the compliance test is failed. You can select Override to enter
Standby mode.You can select Retry to perform automatic circuit leak test again.
FIGURE 3-11 Compliance Test: Fail
The following screen is displayed if the automatic circuit leak test is completed. Select Continue to
enter Standby mode.
FIGURE 3-12 Automatic Circuit Leak Test Completed