System Calibration Calibration
4 - 28 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 4-34
• The screen shown below is displayed after a successful flow sensor calibration. Select Done to
exit the calibration screen.
FIGURE 4-35
16. The screen shown below is displayed if the ongoing calibration is canceled. Select Try Again to
do the calibration again. Select Done to exit the calibration screen.