A5/A3™ Service Manual 046-001141-00 4 - 35
Calibration System Calibration
FIGURE 4-46
12.Select Cancel and the screen shown below is displayed.
FIGURE 4-47
NOTE: After flow calibration, check the accuracy of flow sensors by referring to
Section 3.8.6 Check the Flow Sensor Accuracy.
NOTE: In case of calibration failure, first fix the problem and then perform flow
calibration again.