A5/A3™ Service Manual 046-001141-00 4 - 53
Calibration System Calibration
FIGURE 4-67
9. Select Cancel and the screen shown below is displayed.
FIGURE 4-68
NOTE: After pressure calibration, test the accuracy of pressure sensors by
referring to 3.7.5 Check the Pressure Sensor Accuracy.
NOTE: In case of calibration failure, first fix the problem and then perform
pressure calibration again.