System Calibration Calibration
4 - 68 046-001141-00 A5/A3â„¢ Service Manual
FIGURE 4-85
FIGURE 4-86
5. Turn the O2, AIR and N2O needle valves to the maximum positions. Turn the AIR resistor on the
pneumatic resistor block assembly to cause the AIR flow to fall within 15 to 15.5 L/min.
N2O Resistor