System Calibration Calibration
4 - 72 046-001141-00 A5/A3™ Service Manual
For O2 and AIR, set the output pressure using the table below. For N2O, set the output pressure to 58
psi.
1. Turn off the power supply and all gas supplies. Open the service door.
FIGURE 4-91
2. Remove the sheet metal at the rear of the equipment.
cylinder pressure (psi) regulator pressure (psi)
1000 52.2
1250 50.6
1500 49
1750 47.4
2000 47.1
2250 46.8