Technical Alarms Check Repair and Troubleshooting
5 - 16 046-001141-00 A5/A3™ Service Manual
EtDES Overrange
Low
Low
The monitored value 
exceeded the module 
measurement range.
1. Reduce the concentration of the monitored gas to the 
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.
FiDES Overrange
Low
Low
The monitored value 
exceeded the module 
measurement range.
1. Reduce the concentration of the monitored gas to the 
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.
Rate Overrange Low
The monitored value 
exceeded the module 
measurement range.
1. Reduce the concentration of the monitored gas to the 
normal range.
2. Re-calibrate the AG module.
3. Replace the AG module.