A5/A3â„¢ Service Manual 046-001141-00 6 - 35
Repair and Disassembly Disassemble the Assemblies
3. Remove the two screws on the needle valve assembly to remove the assembly.
FIGURE 6-63
6.2.4.6 Remove the Touch Screen
1. Remove display assembly.
2. Unplug the related cables from the touch screen control board.
FIGURE 6-64