Calibrate the AG Module Anesthetic Gases and O
2
Concentration Monitoring
7 - 12 Operator’s Manual of Anesthesia System
2. Ensure that the system is in Standby mode. Otherwise, select the [End Case] soft key on the
Main Screen and follow the on-screen prompts to enter the standby mode.
3. Select the soft key > [System] soft key (system password required).
4. Select the [Calibrate] soft key.
5. Select the [AG Module] or [Internal AG Module] soft key.
6. Wait for the AG module to be fully warmed up.
7. Enter the actual concentration of the calibration gas.
8. Turn on the calibration gas canister. The system displays the real-time concentration of
calibration gas.
9. Select the [Begin] soft key to start calibrating the AG module. The system will display the
results of the calibration status when the process is completed.
10. Select the [Done] soft key after the calibration is successful. If the calibration failed, select
the [Retry] soft key to calibrate the module again.
11. Select to close the window.