Response Time of the Anesthesia System to an Increase in O2 Concentration (From 21% to 90%/100%) Product Specifications
10 - 26 Operator’s Manual of Anesthesia System
10.14 Response Time of the Anesthesia System to
an Increase in O
2
Concentration (From 21%
to 90%/100%)
Ventilator setting: TV=500 ml, Freq=10 bpm, I:E=1:2
Apnea alarm delay time is adjustable, Range: 5s to 60s, Accuracy: ±3s
CO2 apnea delay time is adjustable, Range: 10s to 40s, Accuracy: ±3s
It has Negative Pressure alarm function, when Negative Pressure in Airway alarm is triggered, Paw shall
be less than -10cmH2O.
TABLE 10-39
Alarm Limits
Fresh Gas(L/min) O
2
Flush Response Time
From 21% to 90%
Response Time
From 21% to 100%
1.0 Without O
2
Flush 10 minutes 20 minutes
With O
2
Flush 30 seconds 1 minute
5.0 Without O
2
Flush 1 minute 30 seconds 4 minutes
With O
2
Flush 30 seconds 1 minute
10.0 Without O
2
Flush 30 seconds 1 minute
With O
2
Flush 30 seconds 1 minute
TABLE 10-40 Response Time of the Anesthesia System to an Increase in O
2
Concentration (From
21% to 90%/100%)

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals