Environmental Specifications Product Specifications
10 - 6 Operator’s Manual of Anesthesia System
10.4 Environmental Specifications
10.5 Electrical Specifications
10.5.1 Main Electrical Power Specifications
The anesthesia system complies with ANSI/AAMI ES60601-1 for its main power supply.
Operating Temperature: +10°C to +40°C
+50°F to 104°F
Storage Temperature: -20°C to +60°C (Oxygen Cell: -20°C to 50°C)
-4°F to 140°F (Oxygen Cell: -4°F to 122°F )
Operating Humidity: 15 to 95% RH, non-condensing
Storage Humidity: 10 to 95% RH, non-condensing
Operating Atmospheric Pressure: 70 kPa to 106.7 kPa (10 psi to 15.5 psi)
Storage Atmospheric Pressure: 50 kPa to 106.7 kPa (7.3 psi to 15.5 psi)
TABLE 10-4 Environmental Specifications
Power Supply Input Voltage: 100 to 120 VAC @ 50/60 Hz
Power Supply Input Current: 10 A maximum (5 A maximum for anesthesia system, 5A
maximum for auxiliary outlets)
Length of Power Cord: 5 m
Grade of Power Cord: Hospital grade
TABLE 10-5 Main Electrical Power Specifications