Safety Information Safety
1 - 10 Operator’s Manual of Anesthesia System
CAUTION:  Turn the flow control knob of the backup flow control system slowly. 
To avoid damaging the control valves, do not turn further when the 
flowmeter reading is out of range. When turning a flow control knob 
clockwise to decrease flow, the flowmeter should reach 1 L/min before 
the knob reaches its most clockwise mechanical stop (off) position. 
Do not turn any further when the knob has reached the off position. 
Turning a flow control knob counter clockwise increases flow.
CAUTION:  Prevent or avoid using and storing the gas supply hose assembly in an 
environment exposed to ultraviolet light or oxidizing agents, or in a 
high-temperature or moist environment to avoid damage to people 
and materials because of the release of pressure from aged hoses in the 
assembly.
1.1.3 Notes
NOTE:  Figures in this manual are provided for reference purposes only. 
Screens may differ based on the system configuration and selected 
parameters.
NOTE:  Put the equipment in a location where you can easily see the screen and 
access the operating controls.
NOTE:  Keep this manual close to the equipment so that it can be obtained 
conveniently when needed.
NOTE:  The software was developed in compliance with IEC 60601-1.The 
possibility of hazards arising from software errors is minimized.
NOTE:  This manual describes all features and options. Your equipment may 
not have all of them.
NOTE:  The equipment is intended to be operated with its integral Breathing 
Pressure monitoring in use.
NOTE:  The equipment is intended to be operated with its integral Breathing 
Pressure limiting devices in use.
NOTE:  The equipment is intended to be operated with its integral Expiratory 
Volume monitoring in use.
NOTE:  The equipment is intended to be operated with its integral Breathing 
System integrity Alarm System in use.
NOTE:  The equipment is intended to be operated with its integral Continuous 
Pressure Alarm in use.
NOTE:  The equipment is intended to be operated with its integral O
2
 
monitoring in use.
NOTE:  An Anesthesia Vapor Delivery Device is to be used with an Anesthetic 
Agent Monitor complying with ISO 80601-2-55. The connection of 
Patient Circuit and Agent monitor should be made by a sample line.
NOTE:  Continuously monitor the anesthetic agent concentration when using the 
anesthesia system to ensure accurate output of the anesthetic agent.
NOTE:  Check the liquid level of the anesthetic agent before and during all 
operations. When the liquid level is below the warning line, more 
anesthetic agent needs to be added. Refer to the vaporizer Instructions 
For Use for filling the vaporizer and other information.