Spirometry Screen Parameters and Factory Defaults
B - 2 Operator’s Manual of Anesthesia System
B.1 Spirometry Screen
B.2 Limits
ITEM DEFAULT
Loop Type P-V loop, F-V loop
Save Loop Baseline loop
Show Reference Off
Review Loops P-V loop
PARAMETERS DEFAULT
Alarms: Limits: PEAK High Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Neonate: 40 cmH
2
O
Alarms: Limits: PEAK Low Adult: 10 cmH
2
O
Pediatric: 8 cmH
2
O
Neonate: 8 cmH
2
O
Alarms: Limits: MV High Adult: 12 L/min
Pediatric: 6 L/min
Neonate: 6 L/min
Alarms: Limits: MV Low Adult: 1 L/min
Pediatric: 1 L/min
Neonate: 0.2 L/min
Alarms: Limits: Vte High 1000 ml
Alarms: Limits: Vte Low Off
Alarms: Limits: RR High Off
Alarms: Limits: RR Low Off
Alarms: Limits: EtCO
2
High Adult or Pediatric: 50 mmHg
Neonate: 45 mmHg
Alarms: Limits: EtCO
2
Low Adult or Pediatric: 25 mmHg
Neonate: 30 mmHg
Alarms: Limits: FiCO
2
High 4 mmHg
Alarms: Limits: Apnea Delay Time 30 sec
Alarms: Limits: CO
2
Apnea Delay Time 30 sec
Alarms: Limits: FiO
2
High Off
Alarms: Limits: FiO
2
Low 18 %
Alarms: Limits: EtO
2
High 88 %
Alarms: Limits: EtO
2
Low Off
Alarms: Limits: FiN
2
O High 53 %
Alarms: Limits: FiN
2
O Low 0 %
Alarms: Limits: EtN
2
O High 55 %
Alarms: Limits: EtN
2
O Low 0 %
Alarms: Limits: AGSS Alarms On
Alarms: Agents: EtHal High 3 %
Alarms: Agents: EtHal Low 0 %
Alarms: Agents: FiHal High 2 %