Alarms and Prompt Messages Alarms and Messages
8 - 28 Operator’s Manual of Anesthesia System
Incompatible
AG Software
Version
Incompatible version
found for the AG
module.
Replace the AG
module.
High Runtime No
CO
2
Over
Range
The monitoring value
exceeds the
measurable range.
1. Calibrate the AG
module.
2. Replace the AG
module.
3. Please contact
Mindray Technical
Support if the
problem persists.
Low Runtime Yes
N
2
O Over
Range
HAL Over
Range
ENF Over
Range
ISO Over
Range
SEV Over
Range
DES Over
Range
O
2
Over
Range
awRR Over
Range
Internal AG
Error 01
1. Internal AG
hardware error.
2. Internal AG self-test
error.
3. Internal AG
hardware fault.
4. Internal AG
initialization error.
5. Internal AG
communication
stopped.
Please contact
Mindray Technical
Support.
Low Runtime Yes
Internal AG
Error 02
Internal AG Zero
Failed.
Internal AG
Error 03
Internal AG No
Water trap.
Internal AG
Error 04
Internal AG airway
occluded.
Internal AG
Error 05
Internal AG Change
Water trap.
INFORMA-
TION CAUSE ACTION
ALARM
PRIORITY
MACHINE
MODE
WHEN
CHECKED
DISABLED
IN
STANDBY
MODE
Table 8-9 AG Module Alarm Messages

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals