Cleaning and Disinfection Maintenance
9 - 18 Operator’s Manual of Anesthesia System
Figure 9-18 Put out the volume exchanger
3. Close the volume exchanger cover.
9.7.3.1.11 Visual Inspection
After the disassembly, inspect the following components for damage, wear and tear, and the visual
inspection criteria is no damage, no creak, no distortion.
•Volume Exchanger
• Breathing System Body
• Bag Arm
•O
2
sensor support
•Bypass
•CO
2
Absorbent Canister
• Inspiration connector
• Expiration connector
• Flow sensors
•Inspiration valve
• Expiration valve
• Seals and sealing rings
If the components are damaged, worn and torn, contact your service personnel or Mindray.
9.7.3.2 Cleaning and Disinfection (Choose One)
9.7.3.2.1 Manual Cleaning
1. Submerge or rinse the Breathing System in an authorized cleaning agent (See Table 9-3 "Manual
Cleaning Agent" on Pages 9-9) for the required period as specified by the manufacturer of the
cleaning solution.
2. Rinse the Breathing System with the clean water.
3. Completely dry the Breathing System before use.
4. To disinfect the Breathing System, proceed to following steps. Otherwise, jump to 9.7.3.4
Reassembly.
Volume exchanger

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals