Monitor Module Product Specifications
10 - 16 Operator’s Manual of Anesthesia System
Resolution: CO
2
: 0.1 %
O
2
: 1 %
N
2
O : 1 %
DES : 0.1 %
SEV : 0.1 %
ENF : 0.1 %
ISO : 0.1 %
HAL : 0.1 %
ISO accuracy mode Add ± 0.3 %
ABS
to full accuracy for CO
2
;
Add ± 8 %
REL
to full accuracy for all agents;
N
2
O accuracy is ± (8 %
REL
+ 2 %
ABS
).
Full accuracy mode Gas Range (%) Accuracy (vol.%)
CO
2
0 to 1 ± 0.1
1 to 5 (not including 1) ± 0.2
5 to 7 (not including 5) ± 0.3
7 to 10 (not including 7) ± 0.5
> 10 Unspecified
N
2
O0 to 20 ± 2
20 to 100 (not including
20)
± 3
O
2
0 to 25 ± 1
25 to 80 (not including
25)
± 2
80 to 100 (not including
80)
± 3
DES 0 to 1 ± 0.15
1 to 5 (not including 1) ± 0.2
5 to 10 (not including 5) ± 0.4
10 to 15 (not including
10)
± 0.6
15 to 18 (not including
15)
± 1
> 18 Unspecified
SEV 0 to 1 ± 0.15
1 to 5 (not including 1) ± 0.2
5 to 8 (not including 5) ± 0.4
> 8 Unspecified
ENF, ISO, HAL 0 to 1 ± 0.15
1 to 5 (not including 1) ± 0.2
> 5 Unspecified
TABLE 10-23 AG Module