Operator’s Manual of Anesthesia System 10 - 19
Product Specifications Monitor Module
10.10.1.2 Effect of Interfering Gas on AG Measured Value
EtO
2
High Limit Off, 12 to 100 1 %
EtO
2
Low Limit Off, 10 to 98
FiO
2
High Limit Off, 20 to 100
FiO
2
Low Limit 18 to 98
EtN
2
O High Limit Off, 2 to 100 1 %
EtN
2
O Low Limit Off, 0 to 98
FiN
2
O High Limit Off, 2 to 100
FiN
2
O Low Limit Off, 0 to 98
EtHal High Limit Off, 0.2 to 5.0 0.1 %
EtHal Low Limit Off, 0.0 to 4.8
FiHal High Limit Off, 0.2 to 5.0
FiHal Low Limit Off, 0.0 to 4.8
EtEnf High Limit Off, 0.2 to 5.0 0.1 %
EtEnf Low Limit Off, 0.0 to 4.8
FiEnf High Limit Off, 0.2 to 5.0
FiEnf Low Limit Off, 0.0 to 4.8
EtIso High Limit Off, 0.2 to 5.0 0.1 %
EtIso Low Limit Off, 0.0 to 4.8
FiIso High Limit Off, 0.2 to 5.0
FiIso Low Limit Off, 0.0 to 4.8
EtSev High Limit Off, 0.2 to 8.0 0.1 %
EtSev Low Limit Off, 0.0 to 7.8
FiSevHigh Limit Off, 0.2 to 8.0
FiSev Low Limit Off, 0.0 to 7.8
EtDes High Limit Off, 0.2 to 18.0 0.1 %
EtDes Low Limit Off, 0.0 to 17.8
FiDes High Limit Off, 0.2 to 18.0
FiDes Low Limit Off, 0.0 to 17.8
TABLE 10-24
Alarms
Gas Under
Tes t
Concentration Quantitative Effect (Volume Fraction)
CO
2
N
2
O HAL SEV ISO ENF DES O
2
N
2
O 60 % 0.1% / 0.1% 0.1% 0.1% 0.1% 0.1% 0.2%
HAL 4 % 0.1% 0.1% / 0.1% 0.1% 0.1% 0.1% 1.0%
SEV 5 % 0.1% 0.1% 0.1% / 0.1% 0.1% 0.1% 1.0%
ISO 5 % 0.1% 0.1% 0.1% 0.1% / 0.1% 0.1% 1.0%
ENF 5 % 0.1% 0.1% 0.1% 0.1% 0.1% / 0.1% 1.0%
DES 15 % 0.1% 0.1% 0.1% 0.1% 0.1% 0.1% / 1.0%
NOTE: Test GAS LEVELS shall be ±20% of the specified level.
TABLE 10-25 Effect of Interfering Gas on AG Measured Value