Safety Information Safety
1 - 8 Operator’s Manual of Anesthesia System
CAUTION: Before clinical use, the equipment must be correctly calibrated and/or
the respective tests must be performed, as described in this Operators
Manual.
CAUTION: If system faults occur during the initial calibration or testing, the
equipment should not be operated until those faults have been
corrected by a qualified service personnel.
CAUTION: After servicing, functional, sensor, and system tests must be performed
before clinical use.
CAUTION: Only vaporizers with Selectatec Interlock-Systems may be used with
this equipment.
CAUTION: Each time you replace the vaporizer, please carry out leak test for the
breathing circuit.
CAUTION: Use cleaning agent sparingly. Excess fluid could enter the equipment
and cause damage.
CAUTION: Do not autoclave any parts of the equipment unless specifically
identified as autoclavable in this manual. Clean the equipment only as
specified in this manual.
CAUTION: To prevent system damage:
Refer to the documentations provided by the manufacturer of the
cleaning agent.
Never use organic, halogenated or petroleum-based solvents,
anesthetic agents, glass cleaning agents, acetone or other irritant
agents.
Never use abrasive agents (i.e. steel wool or silver polish) to clean
components.
Keep all liquids away from electronic components.
Prevent liquid from entering the equipment.
All cleaning solutions used must have a pH value between
7.0 and 10.5.
CAUTION: Never immerse the oxygen sensor or its connector into any type of
liquid. Dispose the O
2
sensor according to the manufacturer's
specifications.
CAUTION: Do not fumigate using peracetic acid or formaldehyde.
CAUTION: The valve disc in each of the inhalation and exhalation valve assemblies
on the breathing system is fragile and must be handled with care while
disassembling the valve cage from the valve assembly.
CAUTION: Only connect Mindray approved devices to the equipment's
communication ports. Devices connected to the ethernet ports must
comply with IEC 60950.
CAUTION: Do not connect any non-isolated devices to the DB9/RS232C interface
of the equipment.
CAUTION: Do not connect any devices to the SB ports other than Mindray
approved USB storage devices and a supported USB mouse.
CAUTION: Do not wash the inner surface of the oxygen sensor.

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals