Operator’s Manual of Anesthesia System B - 9
Parameters and Factory Defaults Ventilation Mode
PEEP 0 cmH
2
O0 cmH
2
O0 cmH
2
O0 cmH
2
O0 cmH
2
O0 cmH
2
O0 cmH
2
O—
ΔPsupp —Adult:
8 cmH
2
O
Pediatric:
5 cmH
2
O
Neonate:
5 cmH
2
O
—— Adult:
8 cmH
2
O
Pediatric:
5 cmH
2
O
Neonate:
5 cmH
2
O
Adult:
8 cmH
2
O
Pediatric:
5 cmH
2
O
Neonate:
5 cmH
2
O
Adult:
8 cmH
2
O
Pediatric:
5 cmH
2
O
Neonate:
5 cmH
2
O
——
F- Trig —Adult:
3 L/min
Pediatric:
2 L/min
Neonate:
2 L/min
—— Adult:
3 L/min
Pediatric:
2 L/min
Neonate:
2 L/min
Adult:
3 L/min
Pediatric:
2 L/min
Neonate:
2 L/min
Adult:
3 L/min
Pediatric:
2 L/min
Neonate:
2 L/min
——
P-Trig —-3 cmH
2
O— -3 cmH
2
O-3 cmH
2
O-3 cmH
2
O—
Trig
Window
25% 25% 25%
Tslop e 0.2 s 0.2 s 0.2 s 0.2 s 0.2 s 0.2 s 0.2 s
Exp% 25 % 25 % 25 % 25 %
ΔPapnea —— Adult:
15 cmH
2
O
Pediatric:
10 cmH
2
O
Neonate:
10 cmH
2
O
——
Apnea I:E —— 1:2
Apnea Ti —— Adult: 2.5 s
Pediatric: 1.5 s
Neonate: 1.0 s
——
Phigh —— 15 cmH
2
O—
Plow —— 5 cmH
2
O—
Thigh Adult: 2.5 s
Pediatric: 1.3 s
Neonate: 1.0 s
Tlow Adult: 5.0 s
Pediatric: 2.7s
Neonate: 2.0 s
Alarm —— On
CO
2
Alarms —— On
PARAME-
TERS VCV SIMV-VC PCV PCV-VG SIMV-PC CPAP/PS SIMV-VG APRV MANUAL

Table of Contents

Other manuals for Mindray A8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A8 and is the answer not in the manual?

Mindray A8 Specifications

General IconGeneral
BrandMindray
ModelA8
CategoryMedical Equipment
LanguageEnglish

Related product manuals