EasyManua.ls Logo

Mindray A8 - Page 257

Mindray A8
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual of Anesthesia System B - 11
Parameters and Factory Defaults Ventilation Linkage Parameters
PCV Pinsp PLAT
or 80% of
PEAK or
the closest
approxima-
tion
#— PLAT
or the
closest
approxi-
mation
*# ##
RR ****#*#
I:E *#*####
Tslope #******
PEEP ******#
PCV-VG Vt **Vti mea-
surement or
the closest
approxima-
tion
—## * #
RR *** *#*#
I:E *#* ####
Tslope #** ****
PEEP *** ***#
Plimit **# ##*#
CURRENT VENTILATION
MODE & AFFECTED
PARAMETERS
PREVIOUS VENTILATION MODE
VCV SIMV-VC PCV PCV-VG SIMV-PC CPAP/PS SIMV-VG APRV
* In the event that the previous ventilation mode and the current one share some parameter values, the parameter setting remains unchanged.
# For the parameters that are not shared by the previous ventilation mode and the current one, the most recent settings of the parameters apply.