Electromagnetic Compatibility
E - 8 Operator’s Manual of Anesthesia System
3. Non-delivery of
excessive
concentrations of a
volatile anaesthetic
agent or generation of
a technical alarm
condition
3.1 Delivered
vapour
concentration
//
3.2 Anaesthetic
agent monitoring
equipment
3.2.1 Anaesthetic
agent monitoring
equipment
Accuracy (%):
CO
2
: 0 ± 0.1 vol.%
N
2
O: 0 ± 2 vol.%
O
2
: 21 ± 1 vol.%
DES: 0 ± 0.15 vol.% (only
applicable for equipping with
Desflurane electrical vaporizer)
4. Airway pressure
monitoring and
associated alarm
4.1 Airway
pressure
monitoring
equipment
//
5. Measurement
accuracy and gas
reading alarm
condition or
generation of a
technical alarm
condition (AG module)
5.1 Measurement
accuracy
//
5.2 Alarm
condition priority
//
5.3 Supply failure
technical alarm
condition
//
ANESTHESIA SYSTEM'S ESSENTIAL
PERFORMANCE
ESSENTIAL
PERFORMANCE
TESTED DURING
EMC IMMUNITY
TESTS
CRITERIA DURING EMC
IMMUNITY
TABLE E-6
Anesthesia system's EMC essential performance

Table of Contents