Operating Your Analyzer
6-27
3. CLICK “Enter”and a window will pop up to instruct you how to shut down the analyzer, as
Figure6-21shows;
Figure6-21 Shutdown window
z The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
z The reagents are irritating to eyes, skin and diaphragm. Wear proper
personal protective equipment (e.g. gloves, lab coat, etc.) and follow safe
laboratory procedures when handling them in the laboratory.
4. Present the E-Z cleanser to the sample probe and press the aspirate key. The analyzer
will aspirate the E-Z cleanser and automatically clean the fluidic lines and the baths. The
cleaning progress will be displayed on the screen, as Figure 6-22 shows;
Figure 6-22 Shutdown progress bar
5. When the cleaning is finished, place the switch at the back of the analyzer to OFF (O) to
turn off the analyzer;
6. Empty the waste container.