Maintaining Your Analyzer
10-15
10.3.8 Lyse Test
z Samples, controls, calibrators and waste are potentially infectious. Wear
proper personal protective equipment (e.g. gloves, lab coat, etc.) and follow
safe laboratory procedures when handling them in the laboratory.
In case of any abnormal WBC counts or histograms, you can perform the “Lyse Test”
procedure to check whether the lyse can be dispensed properly.
Follow the steps given below to do so:
1. Push the right door latch in the direction indicated in Figure10-15.
Figure10-15
2. Lift up the front panel latch as indicated in Figure10-16 and open the front panel.
Figure10-16