Maintaining Your Analyzer
10-17
5. At the “Maintenance” screen, SELECT “Lyse Test”. Press [ENTER] and the
analyzer will automatically drain the WBC bath and then dispense 2ml of lyse into
the WBC bath.
6. Check the scale to see whether the lyse has reached the expected line (the first
from bottom).If so, press [ENTER] and the analyzer will automatically flush the bath
and dispense lyse and the test is done.
7. If not, repeat steps 5 and 6 several times. If all the tries have failed, check whether
the lyse has run out or the lyse pickup tube is not properly connected to this
analyzer. If the lyse is still enough and the tube is well connected to the analyzer,
contact the Mindray or your local distributor for repair.
10.3.9 Cleaning Baths
If you suspect the baths are contaminated, follow the steps given below to perform the “Clean
Bath” procedure:
1. At the “Maintenance” screen, SELECT “Clean Baths”.
2. Press [ENTER] to start the procedure.
Figure10-19 Cleaning the baths
3. When the cleaning is done, the screen displays “Clean Baths End”.