Maintaining Your Analyzer
10-35
Figure10-44
6. Loose the retaining screw by a screwdriver, as Figure10-45 shows;
Figure10-45
7. Remove the probe from the wipe block and insert the localizer into the wipe block from
the bottom, as Figure10-46 shows;
Figure10-46
8. Insert the probe into the wipe block until it reaches the localizer, as Figure10-47 shows;