Customizing the Analyzer Software
5-28
5.9.2 Setting the expiration date
ENTER the desired expiration dates into the ” Diluent”, “Rinse” and ” Lyse” boxes.
z For any reagent, the entered expiration date should be either the expiration
date printed on the labeling or the open-container expiration date, whichever
is earlier.
z The open-container expiration date is calculated as follows: the date that
container is opened + the open-container stability days.
5.9.3 Exiting the “ReagentExp. Date” screen
Press [MENU] or [MAIN] to exit the ” ReagentExp. Date” screen and a message box will pop
up to ask you save the changes, as Figure 5-39 shows.
Figure 5-39 A message box to confirm the changes
CLICKEnter” to save the changes and exit to the system menu or the “Count” screen;
CLICKCancel” to abort the changes and exit to the system menu or the “Count
screen.

Table of Contents