5-14 Checking Performance and Functions
4.
In non-frozen mode, a distant target may be similar to a noise dot which should
be discarded.
Image effect is show in figure below:
5.5.2.3 Geometric Positioning Accuracy
Axial Geometric Positioning Accuracy
Test Procedure:
1. Adjusting steps are the same with the Maximum Detection Depth.
2. Record the separation values with measuring caliper in step of 20 mm on the axial target
group.
3. Select all measurement values deviating largely from 20 mm, and calculate the error by the
following formula.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M9 and is the answer not in the manual?

Mindray M9 Specifications

General IconGeneral
TypePortable Ultrasound System
BatteryRechargeable lithium-ion battery
Battery LifeUp to 90 minutes
Display15.6 inch high resolution LED monitor
Imaging ModesB, M, Color, Power, PW, CW
TransducersLinear, Convex, Phased Array
Transducer Ports3
Probe Connectors3
ConnectivityDICOM, USB, HDMI
Operating SystemWindows-based