System Diagnosis and Support 10-3
10.2 Get Whole Machine Started
Power-on to the
machine
User interface
Press power button on
control panel
Temporary black screen
Windows start-up
graphics appears
Temporary black screen.
Doppler start-up graphics
appears (including progress
bar)
Windows operating system
starts
Doppler starts
Power-on completes. BIOS
starts.
Power-on indicator is
on
Hard disk indicator
blinks
Some mode buttons become
orange (B mode button)
1.Backlight of control
panel is on.
2.Power status indicators
of IO back board are on.
3.The fan starts running.
BIOS start-up
graphics appears
Windows logging
graphics appears

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M9 and is the answer not in the manual?

Mindray M9 Specifications

General IconGeneral
TypePortable Ultrasound System
BatteryRechargeable lithium-ion battery
Battery LifeUp to 90 minutes
Display15.6 inch high resolution LED monitor
Imaging ModesB, M, Color, Power, PW, CW
TransducersLinear, Convex, Phased Array
Transducer Ports3
Probe Connectors3
ConnectivityDICOM, USB, HDMI
Operating SystemWindows-based