Structure and Assembly/Disassembly 8-41
4. Remove holder bracket, hook base of probe cable, moveable cable hook.
Unscrew 6 screws (M4 X 12 cross panhead screw, 3 screws for each left and right side) on the
holder bracket, and then take out the holder bracket.
Unscrew 2 screws (M4 X 12 cross panhead screw) on the base of probe holder, and take out
the base of probe holder.
Unscrew 2 screws (M4 X 12 cross panhead screw, 1 screw for each left and right side) on the
cable hook, and then take out the cable hook.
M4X12 cross panhead
screw (10 screws)
Cable hook (one
for each left and
right side)
Base of
probe holder
Holder bracket
(one for each left
and right side)
M4X12cross panhead screw (4 screws; 2
screws for each left and right side)