6-23 
 
The first three cycles of AMV is PCV experimental ventilation to calculate patient’s lung 
resistance and compliance. Initial ventilation parameters are: 
Adult experimental ventilation cycle setting parameters 
IBW (kg)  Pinsp(cmH
2
O) Tinsp(s) f(/min) 
10-29 15  1  15 
30-39 15  1  14 
40-59 15  1  12 
60-89 15  1  10 
90-99 18  1.5  10 
≥100 20  1.5  10 
Pediatric experimental ventilation cycle setting parameters 
IBW (kg)  Pinsp(cmH
2
O) Tinsp(s) f(/min) 
3-5 15  0.4  30 
6-8 15  0.6  25 
9-11 15  0.6  20 
12-14 15  0.7  20 
15-20 15  0.8  20 
21-23 15  0.9  15 
24-29 15  1  15 
30-35 15  1  14 
After three experimental ventilation, enter the automatic adjustment stage. Based on the 
principle of minimum WOB, ensure that the actual minute volume is as close as possible to 
the preset minute volume value. Mandatory ventilation is administered if the patient has no 
spontaneous breathing. Support ventilation is administered if the patient restores spontaneous 
breathing. 
The following figure shows typical waveforms in AMV mode. 
Pressu re
Flow
Time
Time
Test 
ventilation
Pressure 
support level
Exp%
Inspiratory 
trigger icon
Patients without 
spontaneous patient 
breathing
Mechanically 
operated ventilation
Patients with 
spontaneous breathing 
Pressure support
Mechanically 
operated pressure