Mindray SV300 - Factory Default Alarm Settings

Mindray SV300
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-7
E.6 Alarm
Alarm Factory default setting
Paw high alarm limit 50 cmH
2
O
MV high alarm limit Adult: 7.4 L/min; pediatric: 3.2 L/min
MV low alarm limit Adult: 2.9 L/min; pediatric: 1.3 L/min
TVe high alarm limit Adult: 980 mL; pediatric: 210 mL
TVe low alarm limit Adult: 245 mL; pediatric: 55 mL
FiO
2
high alarm limit 100 %
FiO
2
low alarm limit 21 %
EtCO
2
low alarm limit Adult: 15 mmHg; pediatric: 20 mmHg
EtCO
2
high alarm limit 50 mmHg
SpO
2
high alarm limit 100 %
SpO
2
low alarm limit 90 %
Desat alarm limit 80 %
PR high alarm limit Adult: 120; pediatric: 160
PR low alarm limit Adult: 50; pediatric: 75
f high alarm limit OFF
Tapnea 15 s
Alarm volume 5
E.7 History Data
Trend log Factory default setting
Graphic trend-display group All
Graphic trend-zoom 10 min
Tabular trend-display group All
Tabular trend-interval 1 min
Event logbook-filter All Events
E.8 Special Functions
Special function Factory default setting
Nebulizer time 30 min
P-V tool-Pstart 3 cmH
2
O
P-V tool-Pmax 15 cmH
2
O
P-V tool-Flow 6 L/min
P-V tool-Vlimit 770 mL
Recruitment Tool-
Pinsp 5 cmH
2
O