10-10
10.10.10 FiO
2
Too High
1. Connect the ventilator to low-pressure O
2
supply. Set the O
2
supply type to LPO.
2. Connect a test lung to the ventilator and start ventilation.
3. Set the FiO
2
high alarm limit to be less than the current O
2
concentration monitored
value after ventilation is stable.
4. Verify that the [FiO
2
Too High] alarm is activated.
10.10.11 FiO
2
Too Low
1. Connect the ventilator to high-pressure O
2
supply. Set the O
2
supply type to HPO.
2. Connect a test lung to the ventilator and start ventilation.
3. Switch off the high-pressure O
2
supply after ventilation is stable.
4. Verify that the [FiO
2
Too Low] alarm is activated.
10.10.12 EtCO
2
Too High
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the CO
2
test module. Select the [Setup] key [Sensor] [CO
2
] to set
[Monitoring] to [ON].
3. After CO
2
warm-up is completed and the CO
2
module enters operating mode, deliver
3 % to 7 % of CO
2
standard gas to the sampling port of sidestream CO
2
module or the
airway adapter of mainstream CO
2
module. Set the EtCO
2
high alarm limit to be less
than the standard gas concentration.
4. Verify that the [EtCO
2
Too High] alarm is activated.
10.10.13 EtCO
2
Too Low
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the CO
2
test module. Select the [Setup] key [Sensor] [CO
2
] to set
[Monitoring] to [ON].
3. After CO
2
warm-up is completed and the CO
2
module enters operating mode, deliver
3 % to 7 % of CO
2
standard gas to the sampling port of sidestream CO
2
module or the
airway adapter of mainstream CO
2
module. Set the EtCO
2
low alarm limit to be greater
than the standard gas concentration.
4. Verify that the [EtCO
2
Too Low] alarm is activated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min