7-6
7.2.2.3 Set BTPS Compensation
The CO
2
module is configured to compensate CO
2
readings for either Body Temperature and
Pressure, Saturated Gas (BTPS), to account for humidity in the patient’s breath, or Ambient
Temperature and Pressure, Dry Gas (ATPD).
1. ATPD:
100/%)()(
22 ambco
PvolCOmmHgP
2. BTPS:
100/)47(%)()(
22
ambCO
PvolCOmmHgP
where,
2CO
P
= partial pressure,
%vol
= CO
2
concentration,
amb
P
= ambient pressure, and
unit is mmHg.
For CO
2
module, BTPS compensation is switched on or off based on the actual situations.
1. Select the [Setup] key[Sensor][CO2 Module].
2. Set [BTPS Comp] to [ON] or [OFF] in either BTPS or ATPD.
7.2.2.4 Set Unit
1. Select the [Setup] key[Maintain][User]enter the required password[Setting]
[Unit].
2. Set [CO
2
Unit] to [mmHg], [kPa] or [%].
7.2.2.5 Set CO
2
Waveform
Refer to 5.1.1 to set CO
2
waveform.
7.2.3 Measurement Limitations
Measurement accuracy may be compromised due to:
Leakage or internal leakage of the sample gas
Mechanical shock
Cyclic pressure which is greater than 10 kPa (100 cmH
2
O)
Other interference source (if available)
Measurement accuracy may be affected by the breath rate and I/E ratio as follow:
EtCO
2
is within specification for breath rate 60 bpm and I/E ratio 1:1;
EtCO
2
is within specification for breath rate 30 bpm and I/E ratio 2:1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min