B-1
B Product Specifications
The ventilator is already integrated with expiratory volume monitor, pressure measurement
device, and pressure release device. It is equipped with alarm system, O
2
monitor, CO
2
monitor and SpO
2
monitor, where:
The expiratory volume monitor, pressure measurement device, and pressure release
device comply with ISO 80601-2-12 .
The alarm system complies with IEC 60601-1-8.
The O
2
monitor complies with ISO 80601-2-55.
The CO
2
monitor complies with ISO 80601-2-55.
The gas supply hose assembly complies with ISO 5359.
The SpO
2
monitor complies with ISO 80601-2-61.
B.1 Safety Specifications
Type of protection against
electric shock
Class I equipment with internal electrical power supply.
Degree of protection against
electric shock
BF, defibrillation-proof
Operating mode Continuous
Degree of protection against
hazards of explosion
Ordinary equipment, without protection against explosion; not
for use with flammable anaesthetics.
Degree of protection against
harmful ingress of water
Degrees of protection provided by enclosures(IP Code)—IP21
Protection Index according the EN 60529 standard:
2: Protected against solid foreign objects of 12.5 mm diameter
and greater
1: Protected against vertically falling water drops
Electrical connections between
the equipment and the patient
Non-electrical connections
Equipment type Mobile