D-3
CO
2
module
EtCO
2
Too High M
The monitored parameter value exceeds the alarm limit.
1. Check the patient type.
2. Check the alarm limits.
EtCO
2
Too Low M
The monitored parameter value exceeds the alarm limit.
1. Check the patient type.
2. Check the alarm limits.
Apnea CO
2
M
The time of failure to detect respiration by the CO
2
module exceeds Apnea Tinsp.
1. Check the patient.
2. Check apnea time setup.
3. Check the connections of CO
2
module sampling device.
SpO
2
module
SpO
2
Too High M
The monitored parameter value exceeds the alarm limit.
Check the patient’s physiological condition. Check if the
patient type and the alarm limit settings are correct.
SpO
2
Too Low M
The monitored parameter value exceeds the alarm limit.
Check the patient’s physiological condition. Check if the
patient type and the alarm limit settings are correct.
SpO
2
Desat H
The SpO
2
value falls below the desaturation alarm limit.
Check the patient’s condition and check if the alarm limit
settings are correct.
PR Too High M
The monitored parameter value exceeds the alarm limit.
Check the patient’s physiological condition. Check if the
patient type and the alarm limit settings are correct.
PR Too Low M
The monitored parameter value exceeds the alarm limit.
Check the patient’s physiological condition. Check if the
patient type and the alarm limit settings are correct.
No Pulse H
The pulse signal was so weak that the monitor cannot
perform pulse analysis.
Check the patient’s condition, SpO
2
sensor and
measurement site.