D-4
D.2 Technical Alarm Messages
Source Alarm message P Cause and action
Power
board
Battery 1 Failure 01 H
The temperature of battery 1 is higher than expected.
Contact your service personnel.
Battery 1 Failure 02 H
Battery 1 Charge Failure
Contact your service personnel.
Battery 1 Failure 03 H
Battery 1 Aging
Contact your service personnel.
Battery 1 Failure 04 H
Battery 1 Comm Error
Contact your service personnel.
Battery 1 Failure 05 H
Battery 1 Failure
Contact your service personnel.
Battery 2 Failure 01 H
The temperature of battery 2 is higher than expected.
Contact your service personnel.
Battery 2 Failure 02 H
Battery 2 Charge Failure
Contact your service personnel.
Battery 2 Failure 03 H
Battery 2 Aging
Contact your service personnel.
Battery 2 Failure 04 H
Battery 2 Comm Error
Contact your service personnel.
Battery 2 Failure 05 H
Battery 2 Failure
Contact your service personnel.
Battery Temp.
High. Connect
Ext.Pwr.
M
Battery temperature is a bit high during discharge.
Connect to the external power supply.
Battery Temp High.
Syst maybe Down
H
Battery temperature is too high during discharge. The
system may be down.
Connect to the external power supply.
Battery in Use L
The current system is powered by battery. Connect to
the external power supply.
Connect to the external power supply.
Low Battery.
Connect Ext.
Power.
M
The remaining battery power is lower than a threshold.
Connect to the external power supply.
System DOWN.
Connect Ext.
Power.
H
Battery power is depleted. The system will shut down in
a few minutes.
Connect to the external power supply immediately.
Power Board
Comm Stop
H
Power board communication stops.
Contact your service personnel.