D-5
Battery Undetected H
Battery is not available in the current system.
Contact your service personnel.
Main
control
board
Please Reset Date
and Time
L
Button cell is available in the system. But the clock is
powered down and reset.
Re-set the date and time.
Apnea Ventilation
Ended
L
This alarm is given when apnea ventilation ends. There
is no need to process this alarm.
Key Error L
Hardkey or rotary encoder is depressed continuously for
more than 35s.
Contact your service personnel.
Technical Error 01 M
Keyboard Comm Stop. Keys are faulty.
Contact your service personnel.
Technical Error 02 M
Keyboard Selftest Error.
Contact your service personnel.
Device Failure 04 H
Ctrl Module Init Error.
Contact your service personnel.
Device Failure 05 H
Ctrl Module Comm Stop.
Contact your service personnel.
Device Failure 19 H
Power Board Comm Stop.
Contact your service personnel.
Device Failure 20 H
SpO
2
Comm Stop.
Restart the ventilator or contact your service personnel.
Device Failure 21 H
Pressure Sensor Zero Point Error.
Contact your service personnel.
Monitor
board
Technical Error 03 M
Turbine blower Temp Sensor Failure.
Contact your service personnel.
Technical Error 04 M
Buzzer Failure.
Contact your service personnel.
Technical Error 05 M
Atmospheric Pressure Sensor Failure.
Contact your service personnel.
Technical Error 06 M
HEPA Pressure Sensor Failure.
Contact your service personnel.
Technical Error 07 M
3-way Valve Failure.
Contact your service personnel.
Technical Error 08 M
Nebulizer Valve Failure.
Contact your service personnel.
Technical Error 09 M
Insp. Temp Sensor Failure.
Contact your service personnel.
Device Failure 01 H
Power Supply Voltage Error.
Contact your service personnel.
Device Failure 02 H
Memory Error.
Contact your service personnel.