10-11
10.10.14 SpO
2
Too High
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger, set the SpO
2
Desat alarm limit as 0%, set
the SpO
2
low alarm limit as 0% and the SpO
2
high alarm limit 2%.
4. Verify that the [SpO
2
Too High] alarm is activated.
10.10.15 SpO
2
Too Low
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger, set the SpO
2
high alarm limit as 100% and
the SpO
2
low alarm limit 98%.
4. Grasp the wrist with another hand to press the pulse until the SpO
2
reading is below
98 %, and verify that the [SpO
2
Too Low] alarm is activated.
10.10.16 SpO
2
Desat
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger and set the SpO
2
high alarm limit as 100%,
the SpO
2
low alarm limit 98%, and the SpO
2
Desat 98%.
4. Grasp the wrist with another hand to press the pulse until the SpO
2
reading is below
98 %, and verify that the [SpO
2
Too Low] alarm is activated.
10.10.17 PR Too High
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger and set the PR high alarm limit as 15 1/min.
4. Verify that the [PR Too High] alarm is activated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min