10-11
10.10.14 SpO
2
Too High
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger, set the SpO
2
Desat alarm limit as 0%, set
the SpO
2
low alarm limit as 0% and the SpO
2
high alarm limit 2%.
4. Verify that the [SpO
2
Too High] alarm is activated.
10.10.15 SpO
2
Too Low
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger, set the SpO
2
high alarm limit as 100% and
the SpO
2
low alarm limit 98%.
4. Grasp the wrist with another hand to press the pulse until the SpO
2
reading is below
98 %, and verify that the [SpO
2
Too Low] alarm is activated.
10.10.16 SpO
2
Desat
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger and set the SpO
2
high alarm limit as 100%,
the SpO
2
low alarm limit 98%, and the SpO
2
Desat 98%.
4. Grasp the wrist with another hand to press the pulse until the SpO
2
reading is below
98 %, and verify that the [SpO
2
Too Low] alarm is activated.
10.10.17 PR Too High
1. Connect a test lung to the ventilator and start ventilation.
2. Connect the SpO
2
sensor and activate the SpO
2
monitoring function.
3. Connect the SpO
2
sensor to the index finger and set the PR high alarm limit as 15 1/min.
4. Verify that the [PR Too High] alarm is activated.