10-11 
 
10.10.14 SpO
2
 Too High 
1.  Connect a test lung to the ventilator and start ventilation. 
2.  Connect the SpO
2
 sensor and activate the SpO
2
 monitoring function. 
3.  Connect the SpO
2
 sensor to the index finger, set the SpO
2
 Desat alarm limit as 0%, set 
the SpO
2
 low alarm limit as 0% and the SpO
2
 high alarm limit 2%. 
4.  Verify that the [SpO
2
 Too High] alarm is activated. 
 
10.10.15 SpO
2
 Too Low 
1.  Connect a test lung to the ventilator and start ventilation. 
2.  Connect the SpO
2
 sensor and activate the SpO
2
 monitoring function. 
3.  Connect the SpO
2
 sensor to the index finger, set the SpO
2
 high alarm limit as 100% and 
the SpO
2
 low alarm limit 98%. 
4.  Grasp the wrist with another hand to press the pulse until the SpO
2
 reading is below 
98 %, and verify that the [SpO
2
 Too Low] alarm is activated. 
 
10.10.16 SpO
2
 Desat 
1.  Connect a test lung to the ventilator and start ventilation. 
2.  Connect the SpO
2
 sensor and activate the SpO
2
 monitoring function. 
3.  Connect the SpO
2
 sensor to the index finger and set the SpO
2
 high alarm limit as 100%, 
the SpO
2
 low alarm limit 98%, and the SpO
2
 Desat 98%. 
4.  Grasp the wrist with another hand to press the pulse until the SpO
2
 reading is below 
98 %, and verify that the [SpO
2
 Too Low] alarm is activated. 
 
10.10.17 PR Too High 
1.  Connect a test lung to the ventilator and start ventilation. 
2.  Connect the SpO
2
 sensor and activate the SpO
2
 monitoring function. 
3.  Connect the SpO
2
 sensor to the index finger and set the PR high alarm limit as 15 1/min. 
4.  Verify that the [PR Too High] alarm is activated.