6-3
Total selftest results are listed as follows after all selftest items have been completed:
Pass: all selftest items successfully pass the seftest.
Partially Pass: some selftest items fail, but the mechanical ventilation is allowed.
Fail. Ventilation Disabled: some important selftest items fail, but the mechanical
ventilation is not allowed.
High Leakage, Ventilation Disabled: Expiratory flow sensor test, Pressure sensor test,
Expiration valve test, or Safety valve test fails, the mechanical ventilation is not
allowed.
Cancel: some selftest items cancelled and other selftest items have been successfully
passed.
During System Check, the system prompts [Running] on the right side of the current check
item. In this case, if you select [Skip], the system stops the check of this item immediately
and displays [Cancel] as the check result. The check of the next item begins at the same time.
If you select [Stop], the system stops the check of the current item and also the check of the
remaining items immediately, and displays [Cancel] as the check result.
When O
2
sensor test fails, the [O
2
Calibration] button is displayed. Press this button to open
the O
2
calibration menu, and then to calibrate oxygen concentration.
When checks of all items are completed, if you select [Retry], the system starts a new round
of checking. If you select [Exit], the system exits the check and enters Standby status.
6.3 Select Patient
6.3.1 Set Patient Information on the Ventilator
After System Check is completed, select [Continue] to enter Standby status. Then select
patient. If you select [Last Patient], set ventilation type in the accessed screen, and then
select [Start Ventilation]. If you select [New Adult] or [New Pediatric], set gender,
[Height]/[IBW], ventilation type in the accessed screen, and then select [Start Ventilation].