6-2 
 
To enter the System Check screen, 
  The System Check screen is accessed automatically after powering on the system. 
  On the non-standby screen, select the [Standby] button and enter the Standby status 
after your confirmation. Select the [System Check] button in the Standby status to enter 
the System Check screen. 
The system check screen displays the last system check time. Select the [Details] button to 
query the system check information of the ventilator system, including system check items, 
System Check results, and System Check time. 
 
Connect the gas supply and block the Y piece as illustrated. Then select [Continue] to start 
System Check item by item. 
 
System Check items include: 
  Blower test: test the speed of the turbine blower. 
  O
2
 flow sensor test: test the flow sensor in O
2
 limb. 
  Inspiratory flow sensor test: test the inspiration valve and flow sensor. 
  Expiratory flow sensor test: test the expiratory flow sensor. 
  Pressure sensor test: test the pressure sensors at the inspiratory and expiratory ports. 
  Expiration valve test 
  Safety valve test 
  Leakage (mL/min) 
  Compliance (mL/cmH
2
O) 
  Tube resistance (cmH
2
O/L/s) 
  O
2
 sensor test 
 
System Check result can be: 
  Pass: indicates that check of this item is completed and is passed; 
  Fail: indicates that check of this item is completed but is failed; 
  Cancel: indicates that check of this item is cancelled; 
  O
2
 Supply Failure: indicates that O
2
 supply is insufficient when O
2
 sensor test or O
2
 
flow sensor test is being carried out; 
  Monitoring Off: indicates that sensor monitoring function may not be switched on when 
O
2
 sensor test is being carried out.