7-2
For mainstream CO
2
module, besides the above mentioned CO
2
waveform and EtCO
2
monitored parameter, the measurement also provides:
1. V- CO
2
loop
2. Monitored parameters:
Vdaw: airway dead space.
Vdaw/TVe: ratio of airway dead space to tidal volume.
Vtalv: alveolar tidal ventilation.
V’alv: alveolar minute ventilation.
slopeCO
2
: CO
2
rising slope.
V’CO
2
: CO
2
elimination.
VeCO
2
: exhaled CO
2
volume.
ViCO
2
: inspired CO
2
volume.
Some monitored parameters of the mainstream CO
2
module (Vdaw, Vdaw/Tve, Vtalv, V’alv,
slopeCO
2
, V’CO2, VeCO
2
, ViCO
2
) have reference significance only when the patient is in
stable ventilation status. Stable ventilation status refers to the following situations:
Patient is at rest for at least 30 minutes.
Mechanical ventilation parameters (RR, TV, and etc) remain unchanged.
No operations that may affect the patient’s gas exchange or metabolism.
Some monitored parameters of the mainstream CO
2
module may be inaccurate in the
following situations. The affected parameters include Vdaw, Vdaw/Tve, Vtalv, V’alv,
slopeCO
2
, V’CO2, VeCO
2
, ViCO
2
.
System leakage
Patient ventilation condition is unstable
High frequency ventilation (HFV)
Breathing frequency greater than 35/min
Neonate patient
Non-invasive ventilation type
Other circumstance causing wrong CO
2
and Flow measurements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min