EasyManuals Logo

Mindray SV300 User Manual

Mindray SV300
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
7-4
3. After start-up is finished, the [CO
2
Warm-up] message is displayed. The CO
2
module is
in ISO accuracy mode. If you perform CO
2
measurements during warm-up, the
measurement accuracy may be compromised.
4. After warm-up is finished, the CO
2
module enters full accuracy mode.
NOTE
To extend the lifetime of the water trap and CO
2
module, disconnect the water trap
and set the CO
2
monitoring to OFF when CO
2
monitoring is not required.
It takes approximately 2 minutes from powering on the ventilator to reaching the
sidestream CO
2
monitoring performance specified in section B.10 of this manual.
The sidestream CO
2
measurement can be used, with specified accessories, on
intubated and non-intubated, adult and pediatric patients. A sample of the
respiratory gas is drawn from the patient's breathing circuit through an airway
adapter and a gas sampling line.
When dealing with water trap and sampling line, please comply with related
biohazard regulations.
Please don’t block this connector when the sample gas is emitted from CO
2
Module
gas outlet.
CAUTION
The water trap collects water drops condensed in the sampling line, and therefore
it prevents water drops from entering the module. If the collected water reaches a
certain amount, you should drain it to avoid blocking the airway. Dispose of
accumulated fluids in accordance with the hospital policy or your local regulations.
The water trap has a filter preventing bacterium, vapor and patient secretions
from entering the module. After a long-term use, dust or other substances may
compromise the performance of the filter or even block the airway. In this case,
replace the water trap. Replacing the water trap once a month is recommended.
Or, replace the water trap when it is detected leaky, damaged or contaminated.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min